Expand Capability and Decrease Cost |
|
| |
|
|
|
|
|
|
| |
• Acuity-Adaptable Design - Increase Acuity-Level through Platform Configuration.
• Cost Savings - Build Headwall to AIA Minimum Requirements and use the Platform for Mobile “Extras”.
• Patient Outcomes - Increase Nurse-Patient Contact and Empower Patient Mobility.
• Workflow - Increase Staff Efficiency through Cart Consolidation.
• Patient Space - Expand Beyond the Bed throughout the Room & Hospital.
• Consolidate Equipment - Decrease Room & Hallway Clutter, Improve Transport and Mobility.
• Surge Capacity - A Rapid, Cost-Effective System providing ICU-Level Care Anywhere. |
|
| |
|
8 Power Outlets
2 Oxygen Tanks
2 IV Poles
IV Pumps
Suction
Chest Tubes
Drains & Catheters |
Monitor Defibrillator
Computer
Medication Locked Drawers
Procedure Tray
Bins & Baskets
Ventilator
Battery |
Medication
Supply
Computer
Procedure
Equipment
Bedside |
| |
| Compatible with Mounting Systems from… |
 |
|
|
|
| |
| Patient Space - No Longer Defined by the Bed |
 |
| |
|
|
| |
|
|
|
|
| Technical Specifications |
| |
|
Dimensions and Weight
Base Unit (no IV Pole, no Rings)
|
Width 23 ½”
Depth 24 ¼”
Height (to top surface) 34“
Weight 103 lbs |
|
| Maximum Dimensions (IV Pole & Rings) |
Width 25 ½”
Depth 24 ¼”
Height (to top of IV Pole) 73 ¾”
Weight 115 lbs |
|
| Weight Capacity |
Total Equipment Weight Capacity: 147.6 pounds (67 kg)
IV Pole (each) |
|
Equipment: 30 lbs (13.6 kg) - Below marker line
Fluids: 3 liters |
Top Surface: 100 pounds (45.4 kg)
Oxygen Rings(each): 15 pounds (6.8 kg) |
|
|
|
| |
|
Standards and IEC Classifications
|
Class I Equipment – Accessible conductive parts cannot become live in the event of a basic insulation failure because of the way in which they are connected to the protective earth conductor.
Mode of Operation - Continuous
Type B Applied Part
IPX1 - Protected against dripping water |
|
| Electrical Specifications |
Input: 120V AC 60Hz - 10A
Output: 120V AC 60Hz - 10A Maximum from total of all 8 outlets
Circuit Breaker: 10A Lighted(green) Circuit Breaker - External
Retractable Cord: 10 foot, 10A, Hospital Grade(Green Dot)
Outlets: Hospital Grade(Green Dot)
Isolation Transformer: 120VAC, 50/60Hz, 10A; Input = Output.
Grounding: An Equipotentiality post is provided externally on the back of the electrical box for attachment of medical devices that may be located on the Platform. |
|
|
|
| |
|
|
|
|
|
| |
|
|
|
|
|